B-14
Calibration Reference
Model 2750 Multimeter/Switch System Service Manual
:AC:STEP<n>
(CALibration:PROTected:AC:STEP<n>)
Purpose
To program individual AC calibration steps.
Format
:cal:prot:ac:step<n>
Parameters
1
10mV AC at 1kHz calibration step.
2
100mV AC at 1kHZ calibration step.
3
100mV AC at 50kHz calibration step.
4
1V AC at 1kHz calibration step.
5
1V AC at 50kHz calibration step.
6
10V AC at 1kHz calibration step.
7
10V AC at 50kHz calibration step.
8
100V AC at 1kHz calibration step.
9
100V AC at 50khz calibration step.
10
700V AC at 1kHz calibration step.
11
100mA AC at 1kHz calibration step.
12
1A AC at 1kHz calibration step.
13
2A AC at 1kHz calibration step.
Description
The
:AC:STEP
command programs the 13 individual AC calibration
steps; <n> represents the calibration step number. The appropriate sig-
nal must be connected to the instrument when programming each step,
as summarized in the parameters listed above.
Example
:CAL:PROT:AC:STEP7
Program 10V, 50kHz step.
Summary of Contents for 2750
Page 13: ...1 PerformanceVerification...
Page 48: ...1 36 Performance Verification Model 2750 Multimeter Switch System Service Manual...
Page 49: ...2 Calibration...
Page 74: ...2 26 Calibration Model 2750 Multimeter Switch System Service Manual...
Page 75: ...Routine Maintenance 3 RoutineMaintenance...
Page 86: ...3 12 Routine Maintenance Model 2750 Multimeter Switch System Service Manual...
Page 87: ...Troubleshooting 4 Troubleshooting...
Page 111: ...Disassembly 5 Disassembly...
Page 118: ...5 8 Disassembly Model 2750 Multimeter Switch System Service Manual...
Page 119: ...Replaceable Parts 6 ReplaceableParts...
Page 135: ...Specifications A AccuracyandOptimization...
Page 140: ...A 6 Accuracy and Optimization Model 2750 Multimeter Switch System Service Manual...
Page 141: ...Calibration Reference B CalibrationReference...
Page 167: ...Calibration Program C CalibrationProgram...
Page 178: ......
Page 179: ......