Section 3: Basic operations
Model 2290-10 10 kV Power Supply User's Manual
3-4
2290-10-900-01 Rev. A/December 2013
Store and recall
STO (store) and RCL (recall) allow up to nine complete instrument setups to be saved in nonvolatile
memory and later recalled.
To store a setup:
1. Program the various instrument operating modes to be stored.
2. Press the
STO
button.
3. Press a number (
1-9
) to select the desired storage location.
4. Press the
ENTER
button to complete the storage process.
To recall a setup:
1. Press the
RCL
button.
2. Press the number button (
0-9
) for the configuration to be recalled. (RCL 0 returns the instrument
to the factory default setup.)
3. Press the
ENTER
button.
Whenever a setup is recalled, the high voltage is turned off for safety.
If an Err3 (recall error) occurs, the stored setup was lost due to a memory error and must be stored
again using the STO button.
The factory default setup can be recalled by pressing the
CLR
button while turning on the power or by
recalling setup 0 (RCL 0). The default setup is also recalled after a power-on memory error (Err 1).
The next table lists the factory default setup.
Mode
Default setting
Output voltage 0 V
Voltage limit
10000 V
Current limit
1050 µA
Current trip
1050 µA
Reset mode
MAN
High voltage
OFF
GPIB address
14