Model 2290-10 10 kV Power Supply User's Manual
Section 2: Safe configuration and test setup
2290-10-900-01 Rev. A/December 2013
2-3
Connecting the interlock of the Model 2290-10
The interlock pins control the high-voltage supply.
The pin locations and connections are shown in the following figure. The pins are:
•
Pin 3: Chassis ground
•
Pin 2: Interlock enable
•
Pin 1: +5 V DC out (current limited)
Figure 3: Model 2290 interlock connector wiring
The following figure shows how to connect the Model 2290-10 internal 10 V source in order to engage
the external interlock (or interlocks) circuits:
Figure 4: Model 2290-10 Internal 10 kV source