![background image](http://html2.mh-extra.com/html/keithley/2260b-30-36/2260b-30-36_adjustment-manual_3997099035.webp)
Series 2260B Programmable Power Supplies Verification and Adjustment Manual
Section 7: Analog interface adjustment
077104802 / November 2016
7-3
Make sure that you connect the following devices as indicated in the next figure. Also, make sure that
you set the Model 2220-30-1 DC power supply, channel one and channel two to 10.00 V.
Figure 11: Analog adjustment devices
Analog interface adjustment procedure
To start the adjustment procedure:
1. Turn on the programmable DC power supply (Series 2260B) by pressing the Function key.
The Function key lights up, the display indicates "F - 01." Additionally, note that the "x" in the
diagram below indicates that the value is not fixed.
2. Rotate the Voltage knob until the display indicates "F - 00".
3. Press the Current knob to select the numeric indicator.