Series 2260B Programmable Power Supplies Verification and Adjustment Manual
Section 3: Constant voltage verification
077104802 / November 2016
3-5
Constant voltage ripple and noise verification
This test measures the DC output voltage with the (10:1) 50 Ω probe.
Hazardous voltages may be present on all output and guard terminals. To prevent electrical
shock that could cause injury or death, never make or break connections to any of the
instruments used for this verification test while instruments are powered on. Turn off the
2260B instrument from the front panel or disconnect the main power cord from the rear of
the instrument before handling cables. Putting the equipment into an output-off state does
not guarantee that the outputs are powered off if a hardware or software fault occurs.
Precautions must be taken to prevent a shock hazard by surrounding the test device and
any unprotected leads (wiring) with double insulation for up to 800 volts depending on the
power supply that you have, Category I.
Figure 4: 2260B Constant voltage ripple noise verification