Section 5: Verification test record forms
Series 2260B Programmable Power Supplies Verification and Adjustment Manual
5-8
077104802 / November 2016
800 Volt verification record
Print this page and record your verification results that can be used for reference in the future. Make
sure to keep this record with your power supply.
Model
2260B-800-1
2260B-800-2
2260B-800-4
Serial number:
Verified by
Name:
Company/contact:
Year:
Month:
Day:
Environment
Temperature:
Humidity:
Constant voltage tests:
Voltage programming
and measurement
Model
Minimum
specifications
Results
Maximum specifications
Minimum voltage out
All
-400 mV
+400 mV
Measurement accuracy All
-400 mV
+400 mV
Rated voltage out
All
798.8 V
801.2 V
Measurement accuracy All
798.8 V
801.2 V
CV load regulation
Model
Minimum
specifications
Results
Maximum specifications
All
-405 mV
+405 mV
CV line regulation
Model
Minimum
specifications
Results
Maximum specifications
All
-403 mV
+403 mV
CV ripple and noise
Model
Minimum
specifications
Results
Maximum specifications
Peak-to-peak
800-1
N/A
150 mV
800-2
N/A
200 mV
800-4
N/A
200 mV
RMS
800-1
N/A
30 mV
800-2
N/A
30 mV
800-4
N/A
30 mV