
Section 5: Verification test record forms
Series 2260B Programmable Power Supplies Verification and Adjustment Manual
5-2
077104802 / November 2016
30 Volt verification test record
Print this page and record your verification results that can be used for reference in the future. Make
sure to keep this record with your power supply.
Model
2260B-30-36
2260B-30-72
2260B-30-108
Serial number:
Verified by
Name:
Company/contact:
Year:
Month:
Day:
Environment
Temperature:
Humidity:
Constant voltage tests:
Voltage programming
and measurement
Model
Minimum
specifications
Results
Maximum specifications
Minimum voltage out
All
-10.1 mV
+10.1 mV
Measurement accuracy All
-20.1 mV
+20.1 mV
Rated voltage out
All
29.960 V
30.040 V
Measurement accuracy All
29.95 V
30.5 V
CV load regulation
Model
Minimum
specifications
Results
Maximum specifications
All
-20 mV
+20 mV
CV line regulation
Model
Minimum
specifications
Results
Maximum specifications
All
-18 mV
+18 mV
CV ripple and noise
Model
Minimum
specifications
Results
Maximum specifications
Peak-to-peak
30-36
N/A
60 mV
30-72
N/A
80 mV
30-108
N/A
100 mV
RMS
30-36
N/A
7 mV
30-72
N/A
11 mV
30-108
N/A
14 mV