Keithley Instruments, Inc.
Cleveland, Ohio 44139
DRN.
CKD.
APP.
DATE
DATE
DATE
PART NUMBER
LTR
REVISIONS
APP. DATE
FORM 28777A-SBG
BRUNING 40-21 62198-SBG
PURCHASED ITEM
SPECIFICATIONS
2010 Low-Noise Multimeter
HW 9/5/03
Rev. C
HIGH CREST FACTOR ADDITIONAL ERROR
±(% of reading)
7
Crest Factor:
1–2
2–3
3–4
4–5
Additional Uncertainty:
0.05
0.15
0.30
0.40
AC OPERATING CHARACTERISTICS
2
FUNCTION
DIGITS
RDGS./s
RATE
BANDWIDTH
ACV (all ranges)
6
1
⁄
2 3
0.5 (0.4)
SLOW
3 Hz–300 kHz
and
6
1
⁄
2 3
1.4 (1.5)
MED
30 Hz–300 kHz
ACI (all ranges)
6
1
⁄
2 4
4.0 (4.3)
MED
30 Hz–300 kHz
6
1
⁄
2 3
2.2 (2.3)
FAST
300 Hz–300 kHz
6
1
⁄
2 4
35 (30)
FAST
300 Hz–300 kHz
AC SYSTEM SPEEDS
2, 5
FUNCTION/RANGE CHANGE
6
:
4/s.
AUTORANGE TIME:
<3 s.
ASCII READINGS TO RS-232 (19.2K BAUD)
4
:
50/s.
MAX. INTERNAL TRIGGER RATE
4
:
300/s.
MAX. EXTERNAL TRIGGER RATE
4
:
300/s.
ADDITIONAL LOW FREQUENCY ERRORS
±(% of reading)
SLOW
MED
FAST
20 Hz –
30 Hz
0
0.3
—
30 Hz –
50 Hz
0
0
—
50 Hz – 100 Hz
0
0
1.0
100 Hz – 200 Hz
0
0
0.18
200 Hz – 300 Hz
0
0
0.10
>300 Hz
0
0
0
AC GENERAL
INPUT IMPEDANCE:
1M
Ω
±2% paralleled by <100pF.
ACV INPUT PROTECTION:
1000V.
MAXIMUM DCV:
400V on any ACV range.
ACI INPUT PROTECTION:
3A, 250V fuse.
BURDEN VOLTAGE: 1A Range:
<0.35V rms.
3A Range:
<1V rms.
SHUNT RESISTOR:
0.1
Ω
on all ACI ranges.
AC CMRR:
>70dB with 1k
Ω
in LO lead.
MAXIMUM CREST FACTOR:
5 at full scale.
VOLT HERTZ PRODUCT:
≤
8
×
10
7
V·Hz.
OVERRANGE:
120% of range except on 750V and 3A ranges.
AC NOTES
1. Specifications are for SLOW rate and sinewave inputs >5% of range.
2. Speeds are for 60Hz (50Hz) operation using factory default operating
conditions (*RST). Auto zero off, Auto range off, Display off, includes
measurement and binary data transfer out the GPIB.
3. 0.01% of step settling error. Trigger delay = 400ms.
4. Trigger delay = 0.
5. DETector:BANDwidth 300, NPLC = 0.01.
6. Maximum useful limit with trigger delay = 175ms.
7. Applies to non-sine waves >5Hz and <500Hz. (Guaranteed by design for
Crest Factors >4.3)
8. Applies to 0°–18°C and 28°–50°C.
9. For signal levels >2.2A add additional 0.4% to “of reading” uncertainty.
10. Typical uncertainties. Typical represents two sigma or 95% of manufac-
tured units measure < 0.35% of reading and three sigma or 99.7% <
1.06% of reading.
SPEC-2010
A
18127 Released
SZ
1/10/96
B
27373 Revised
SZ
2/21/03
C
28238 Revised
SZ
9/5/03
CAB
1/9/96
DMS
1/9/96
TRUE RMS AC VOLTAGE AND CURRENT CHARACTERISTICS
ACCURACY
1
: ±(% of r % of range), 23°C ±5 °C
VOLTAGE
CALIBRATION
3 Hz–
10 Hz–
20 kHz–
50 kHz–
100 kHz–
RANGE
RESOLUTION
CYCLE
10 Hz
10
20 kHz
50 kHz
100 kHz
300 kHz
100.0000 mV
0.1 µV
1.000000
V
1.0 µV
90 Days
0.35 + 0.03
0.05 + 0.03
0.11 + 0.05
0.60 + 0.08
4 + 0.5
10.00000
V
10 µV
100.0000
V
100 µV
1 Year
0.35 + 0.03
0.06 + 0.03
0.12 + 0.05
0.60 + 0.08
4 + 0.5
750.000
V
1 mV
TEMPERATURE
COEFFICIENT/°C
8
0.035 + 0.003
0.005 + 0.003
0.006 + 0.005
0.01 + 0.006
0.03 + 0.01
CURRENT
CALIBRATION
3 Hz–
10 Hz–
3 kHz-
RANGE
RESOLUTION
CYCLE
10 Hz
3 kHz
5 kHz
1.000000
A
1 µA
90 Day/1 Year
0.30 + 0.04
0.10 + 0.04
0.14 + 0.04
3.00000 A
9
10 µA
90 Day/1 Year
0.35 + 0.06
0.15 + 0.06
0.18 + 0.06
TEMPERATURE
COEFFICIENT/°C
8
0.035 + 0.006
0.015 + 0.006
0.015 + 0.006
Summary of Contents for 2010
Page 14: ...1 Performance Verification 1 Performance Verification...
Page 28: ...2 Calibration 2 Calibration...
Page 53: ...3 Routine Maintenance 3 Routine Maintenance...
Page 56: ...4 Troubleshooting 4 Troubleshoot ing...
Page 74: ...5 Disassembly 5 Disassembly...
Page 83: ...CONFIDENTIAL PROPRIETARY...
Page 84: ......
Page 85: ......
Page 86: ......
Page 87: ...6 Replaceable Parts 6 Replaceable Parts...
Page 98: ...CONFIDENTIAL PROPRIETARY...
Page 99: ...CONFIDENTIAL PROPRIETARY...
Page 100: ......
Page 101: ...A Specifications A Specifications...
Page 107: ...B Calibration CommandReference B Calibration CommandRef erence...