:STEP10 (CALibration:PROTected:DC:STEP10)
Purpose
To program the 1M
Ω
comprehensive calibration step.
Format
:cal:prot:dc:step10 <Cal_resistance>
Parameter
<Cal_resistance>
= 900E3 to 1.1E6 [
Ω
]
Description
:STEP10 programs the 1M
Ω
comprehensive calibration step. The allowable
range of the calibration resistance parameter is from 900E3 to 1.1E6. Use the
1E6 value whenever possible, or the closest possible value.
Example
:CAL:PROT:DC:STEP10
Program 1M
Ω
calibration step.
:STEP11 (CALibration:PROTected:DC:STEP11)
Purpose
To program the 10mA comprehensive calibration step.
Format
:cal:prot:dc:step11 <Cal_current>
Parameter
<Cal_current>
= 9E-3 to 11E-3 [A]
Description
:STEP11 programs the 10mA comprehensive calibration step. The allowable
range of the calibration current parameter is from 9E-3 to 11E-3. Use the
10E-3 value whenever possible for best results.
Example
:CAL:PROT:DC:STEP11 10E-3
Program 10mA step.
:STEP12 (CALibration:PROTected:DC:STEP12)
Purpose
To program the 100mA comprehensive calibration step.
Format
:cal:prot:dc:step12 <Cal_current>
Parameter
<Cal_current>
= 90E-3 to 110E-3 [A]
Description
:STEP12 programs the 100mA comprehensive calibration step. The allow-
able range of the calibration current parameter is from 90E-3 to 110E-3. Use
the 100E-3 value whenever possible for best results.
Example
:CAL:PROT:DC:STEP12 0.1
Program 100mA step.
:STEP13 (CALibration:PROTected:DC:STEP13)
Purpose
To program the 1A comprehensive calibration step.
Format
:cal:prot:dc:step13 <Cal_current>
Parameter
<Cal_current>
= 0.9 to 1.1 [A]
Description
:STEP13 programs the 1A comprehensive calibration step. The allowable
range of the calibration current parameter is from 0.9 to 1.1. Use a value of
1 whenever possible for best results.
Example
:CAL:PROT:DC:STEP13 1
Program 1A step.
Calibration Command Reference
B-11
Summary of Contents for 2010
Page 14: ...1 Performance Verification 1 Performance Verification...
Page 28: ...2 Calibration 2 Calibration...
Page 53: ...3 Routine Maintenance 3 Routine Maintenance...
Page 56: ...4 Troubleshooting 4 Troubleshoot ing...
Page 74: ...5 Disassembly 5 Disassembly...
Page 83: ...CONFIDENTIAL PROPRIETARY...
Page 84: ......
Page 85: ......
Page 86: ......
Page 87: ...6 Replaceable Parts 6 Replaceable Parts...
Page 98: ...CONFIDENTIAL PROPRIETARY...
Page 99: ...CONFIDENTIAL PROPRIETARY...
Page 100: ......
Page 101: ...A Specifications A Specifications...
Page 107: ...B Calibration CommandReference B Calibration CommandRef erence...