![background image](http://html1.mh-extra.com/html/keithley/2000-scan/2000-scan_instruction-manual_661585063.webp)
8.
Disconnect the temporary leads from the multimeter, and connect the output leads from
the scanner card to the rear panel HI and LO INPUT jacks (see
Figure 4-1
).
9.
Close channel 1 as follows;
A.
From normal display, press CHAN. The instrument will display the following:
CHANNEL SELECTION
CLOSE-CHANNEL OPEN-ALL-CHANNELS
B.
Select CLOSE-CHANNEL, then press ENTER. The Model 2001 will display the
following:
ENTER CHAN#01 (1-10)
C.
Using the range and cursor keys, select channel 01, then press ENTER.
D.
Press EXIT to return to normal display.
10.
Note the resistance reading on the Model 2001. Verify that the reading is <2
Ω
.
11.
Repeat steps 9 and 10 for channels 2-9. Verify that the path resistance for each channel
is as follows:
channels 1-10: <2
Ω
.
12.
After completing the tests, turn off the Model 2001 power, remove the scanner card, and
remove all jumper wires (leave HI and LO shorted for each channel if you intend to
perform contact potential tests).
DMM Rear Panel
(Model 2001/2002
shown)
HI
LO
Channel 1
HI
LO
Channel 2-4
Channel 5
HI
LO
Channel 6
HI
LO
Channel 7-9
Channel 10
HI
LO
OUT A
OUT B
HI
LO
2-Pole
4-Pole
Jumper HI and LO
of all 10 Channels
Model 2000-SCAN
Input
HI
Input
LO
Figure 4-1
Connections for path
resistance checks
Service Information
4-5
Summary of Contents for 2000-Scan
Page 12: ......
Page 13: ...1 General Information General Information ...
Page 17: ...2 CardConnections andInstallation CardConnec tionsandInstal lation ...
Page 35: ...3 Operation Operation ...
Page 58: ...3 24 Operation Section 3 Page 24 Tuesday July 8 2003 2 30 PM ...
Page 59: ...4 Service Information Service Information ...
Page 73: ...5 Replaceable Parts Replaceable Parts ...
Page 76: ...5 4 Replaceable Parts ...
Page 77: ......
Page 78: ......
Page 79: ...A Specifications Specifica tions ...
Page 84: ......
Page 85: ......