![background image](http://html1.mh-extra.com/html/keithley/2000-scan/2000-scan_instruction-manual_661585026.webp)
Resistance connections
2-Pole connections
Figure 2-6
shows typical 2-pole resistor test connections. The 2-pole resistance configuration
can be used to test up to ten DUTs.
H
L
H
H
L
H
L
H
L
H
L
H
L
H
L
H
L
H
L
H
L
H
L
CH 1
CH 2
CH 3
CH 4
CH 5
OUT A
CH 6
CH 7
CH 8
CH 9
CH 10
OUT B
To DMM
Input
2000
-
SCAN
Card
Resistors
Under Test
Resistors
Under Test
HI
LO
L
Figure 2-6
Typical connections
for 2-wire resistance
scanning
2-10
Card Connections and Installation
Summary of Contents for 2000-Scan