O p e r a t i n g I n s t r u c t i o n s
|
KBL Tower beautyShaper
9
EXPLOSIVE ATMOSPHERE
DANGER
Explosive atmosphere
Operation in explosive atmospheres can
cause explosions.
• Do not operate the device in an explo-
sive atmosphere.
SALTY AIR - CONTAMINATED AIR
Salty air, contaminated air may lead to
corrosion of the housing and electrical
components.
Damage to the device
possible!
• Do not operate the device in locations
with aggressive environmental conditions
(e.g. air containing chlorine).
• Do not install the device at swimming
pools.
ELECTROMAGNETIC ENVIRONMENT
Do not use near (24 inches or closer) equipment sen-
sitive to electromagnetic interference.
This device complies with International Standards
(as set forth in manufacturer’s declaration following)
for electromagnetic compatibility for medical electri-
cal equipment and/or systems.
These standards are designed to provide reasonable
protection against harmful interference in a typical
medical installation.
However, because of the proliferation of radio-fre-
quency transmitting equipment and other sources of
electrical noise in healthcare and other environments,
it is possible that high levels of such interference due
to close proximity or strength of a source might dis-
rupt the performance of this device. Medical electri-
cal equipment needs special precautions regarding
EMC, and all equipment must be installed and put
into service according to the EMC information speci-
fied in this manual
(chapter 4.1).
DIFFERENCE IN TEMPERATURE
Risk of damage by switching on the
power supply voltage.
When there are great differences in tempera-
ture between the transportation route and the
installation site the device shall not be put into
operation immediately after being set up.
• Before connecting the device to the
power supply, wait at least for 2 hours.
2.10 Commissioning
DANGER
Damaged mains power cable
Risk of death or fatal injury due to electric
shock when touching a damaged mains
power cable.
• Only the manufacturer or specialized
companies may replace the mains
power cable!
Initial commissioning will be carried out by the man-
ufacturer’s own service technicians or by qualified,
trained and authorized service technicians. The
device will be handed over ready for use.
As studio operator it is your responsibility to observe
the electrical regulations applicable at the installation
site.
The device incorporates special preventive measures
with regard to the electromagnetic compatibility and
must be installed and commissioned according to
the EMC instructions obtained from this operating
instructions.
If the device has not been used for a longer period
of time, then it must first be inspected by qualified,
trained and authorized service technicians, before
being put back into operation again.
Ensure no electromagnetic compatibility issues with
the siting
(chapter 4.1).
Summary of Contents for beautyShaper
Page 1: ...OPERATING INSTRUCTIONS BEAUTY SYSTEM beautyShaper Tower ...
Page 2: ......
Page 44: ...Remarks O p e r a t i n g I n s t r u c t i o n s KBL Tower beautyShaper 44 ...
Page 45: ...Remarks O p e r a t i n g I n s t r u c t i o n s KBL Tower beautyShaper 45 ...
Page 46: ...Remarks O p e r a t i n g I n s t r u c t i o n s KBL Tower beautyShaper 46 ...
Page 47: ......
Page 48: ...Ringstraße 24 26 56307 Dernbach Germany fon 49 0 26 89 94 26 0 fax 49 0 26 89 94 26 66 ...