O p e r a t i n g I n s t r u c t i o n s
|
KBL Tower beautyShaper
14
4. Description of the device
4.1 Standards
This device is in conformity with general requirements for basic safety and essential performance of medical
electrical equipment IEC 60601-1:2005 +A1:2012 and CAN/CSA-C22.2 No.601.1-M90) and the standards for
electromagnetic compatibility IEC 60601-1-2:2014 and EN 60601-1-2:2015 (EMC):
EN61000-3-2; EN61000-3-3;
EN60601-1-2; EN55011; EN61000-4-2; EN61000-4-3; EN61000-4-4; EN61000-4-5; EN61000-4-6; EN61000-
4-8; EN61000-4-11.
According to CISPR 11 the electromagnetic emission characteristic of this device is classified in Group1 and
Class A.
The emission characteristics of this device make it suitable for use in industrial areas and hospitals (CISPR 11
class A). If it is used in a residential environment (for which CISPR 11 class B is normally required) this device
might not offer adequate protection to radio-frequency communication services. The studio operator might need
to take mitigation measures, such as relocating or re-orienting the equipment.
See EMC Tables according to IEC 60601-1-2:2014 and EN 60601-1-2:2015.
If, at any point, you determine that the ESSENTIAL PERFORMANCE of the device has been impaired (due to
electro-magnetic interference or other causes), then terminate installation / operation and consult the factory
before continuing.
Table 1
Guidance and Manufacturer’s Declaration - Electromagnetic Emission
The device is intended for use in the electromagnetic environment specified below. The studio operator of the
device should assure that it is used in such an environment.
Emission tests
Compliance
Electromagnetic environment - Guidance
RF emissions
according to CISPR 11
Group 1
The device uses RF energy exclusively for its internal
FUNCTION. Therefore, its HF emissions are very low and
not likely to cause any interference in nearby electronic
equipment.
RF emissions
according to CISPR 11
Class A
The device is intended for use in facilities other than residen-
tial areas, and for facilities that are directly connected to a
public low-voltage supply system that also supplies residen-
tial buildings.
Summary of Contents for beautyShaper
Page 1: ...OPERATING INSTRUCTIONS BEAUTY SYSTEM beautyShaper Tower ...
Page 2: ......
Page 44: ...Remarks O p e r a t i n g I n s t r u c t i o n s KBL Tower beautyShaper 44 ...
Page 45: ...Remarks O p e r a t i n g I n s t r u c t i o n s KBL Tower beautyShaper 45 ...
Page 46: ...Remarks O p e r a t i n g I n s t r u c t i o n s KBL Tower beautyShaper 46 ...
Page 47: ......
Page 48: ...Ringstraße 24 26 56307 Dernbach Germany fon 49 0 26 89 94 26 0 fax 49 0 26 89 94 26 66 ...