O p e r a t i n g I n s t r u c t i o n s
|
KBL Tower beautyShaper
31
CLEANING AND DISINFECTION
WARNING
Risk of Infection.
Risk of Infection by transferring through
skin contact.
• After each session the device must be
cleaned and disinfected with a clean-
ing and disinfectant agent.
• Use only EPA registered quaternary
ammonium germicidal based detergent
solution (follow product label for use-dilu-
tion), such as Lucasol One Step, to clean
and disinfect the device and its compo-
nents.
• For thorough cleaning and disinfection,
the product must be allowed to remain
in contact with the surface being cleaned
and disinfected for a specified amount
of time. This specified amount of time
depends on the cleaner/disinfectant
which is used. Follow the directions for
use of the cleaner/disinfectant. After this
specified application time remove remain-
ing cleaner/disinfectant from the surfaces
of the device. Use a single-pass paper
towel.
• If the device is determined not to be visu-
ally clean at the end of the cleaning step,
the relevant previous cleaning steps must
be repeated so that a visibly soiled device
is not used again. Removal of residual soil
ensures the effectiveness of the disinfec-
tion process and reduces the risk of infec-
tion to users of the device.
ADDITIONAL CLEANING ACTIVITIES
WARNING
These cleaning works must be car-
ried out by qualified, trained and
authorized service technicians.
Every 6 months
Make sure that the inside and outside of the
device is cleaned and free from dust, fluff or objects.
Inadequate cleaning leads to overheat-
ing. As a result, this can lead to damage
the device!
Signs of electrical overstress
Check the cables and connections on signs
of electrical overstress.
• Discoloration
• Deformations
• Melting
• Blisters on the housing
The affected components shall be replaced
immediately! Use only OEM spare parts.
Summary of Contents for beautyShaper
Page 1: ...OPERATING INSTRUCTIONS BEAUTY SYSTEM beautyShaper Tower ...
Page 2: ......
Page 44: ...Remarks O p e r a t i n g I n s t r u c t i o n s KBL Tower beautyShaper 44 ...
Page 45: ...Remarks O p e r a t i n g I n s t r u c t i o n s KBL Tower beautyShaper 45 ...
Page 46: ...Remarks O p e r a t i n g I n s t r u c t i o n s KBL Tower beautyShaper 46 ...
Page 47: ......
Page 48: ...Ringstraße 24 26 56307 Dernbach Germany fon 49 0 26 89 94 26 0 fax 49 0 26 89 94 26 66 ...