Troubleshooting
No Fault Code in Kayden RCMS
No Fault Code in Kayden RCMS
Logic Module Startup Failure
Product Identification Code does not
match installed firmware.
Attempt FACTORY RESET. If operation
not restored, replace the Electronics
Module.
Attempt FACTORY RESET. If operation
not restored, replace the Electronics
Module.
RCMS Fault Code
RCMS Fault Code
Description of Fault
Description of Fault
Course of Action
Course of Action
Fault Code 9
Fault Code 10
Fault Code List
Description
0%
0%
40% - 65%
Thermal Signal
0%
100%
Thermal Signal
0%
100%
No Fault Code in Kayden RCMS
No Fault Code in Kayden RCMS
Security Checksum Error, Eeprom
memory fault.
Electronics Module Failure (special
case with all bargraph LEDs
illuminated).
Attempt FACTORY RESET. If operation is
not restored replace Electronics Module.
Fault may occur as the result of power supply voltage below
specified minimum (brown-out).
Attempt FACTORY RESET. If operation is
not restored replace Electronics Module.
Fault may occur as the result of power supply voltage below
specified minimum (brown-out).
RCMS Fault Code
RCMS Fault Code
Description of Fault
Description of Fault
Course of Action
Course of Action
Fault Code 11
Fault Code 12
0%
75%
40% - 65%
0% - 100%
Thermal Signal
0%
100%
Thermal Signal
0%
100%
RCMS
= Kayden RCM Software