Instructions for use SURGmatic S201 XL Pro - 1.013.7541 SURGmatic S201 L Pro - 1.013.7540
1 User instructions
4 / 28
1 User instructions
Dear User,
Congratulations on purchasing this KaVo quality product. By following the in-
structions below you will be able to work smoothly, economically and safely.
© Copyright by KaVo Dental GmbH
Please direct all questions regarding the product, service and maintenance to
the
KaVo Technical Service:
Toll-free: 1-888-ASK-KAVO (888-275-5286)
Email: [email protected]
Please refer to the serial number of the product in all inquiries!
For repairs, please contact KaVo Repair Service.
For scheduling or if you have any questions, please contact:
KaVo Repair Service
KaVo Dental Technologies, LLC
11727 Fruehauf Drive
Charlotte, NC 28273 USA
Toll-free Direct Customer Service: 1-888-ASK-KAVO (888-275-5286)
Email: [email protected]
www.kavo.com
Target group
The instructions for use are intended for medical professionals, in particular
dentists and office personnel.
The section on Commissioning is also intended for service technicians.
General marks and symbols
Refer to the chapter on Safety/Warning symbol
Important information for users and service technicians
Action request
Sterilization parameters
▪ Sterilizer with triple pre-vacuum:
- at least 3 minutes at 135°C (275 °F)
- Drying time: 16 min.
▪ Sterilizer using the gravity method:
- at least 10 minutes at 135°C (275 °F)
- drying time 30 minutes or alternatively
- at least 30 minutes at 121°C (250 °F)
- Drying time: 15 min.
Thermodisinfectable