Instructions for use SURGmatic S201 XL Pro - 1.013.7541 SURGmatic S201 L Pro - 1.013.7540
5 Operation | 5.1 Inserting the head
5 Operation
5.1 Inserting the head
CAUTION
The SURGmatic S201 L / XL Pro contra-angle handpiece consists of a
shank and a head.
Risk of injury.
▶
It is not permissible to combine the unit with other heads / shanks.
Note
The head of the SURGmatic S201 XL Pro / L Pro contra-angle handpiece may
be taken off the shank only for the non-fixating pre-cleaning.
WARNING
Detachment of the medical device during treatment.
If the head is not properly locked in place, it can become detached during
treatment.
▶
Do not mount or remove the head while it is rotating. Before each treat-
ment, check that the head is firmly seated and that the clamping ring is
tight.
▶
Rotate the clamping ring in the direction of the arrow until it hits the stop
and hold it there.
▶
Insert the head to the stop. Make sure that the catches engage properly.
▶
Rotate the clamping ring in the direction of the arrow (-> close) and tighten
it.
▶
Attach the spray clip.
▶
Make sure that the spray clip is securely affixed.
5.2 Pulling off the head
▶
Pull the spray clip off.
▶
Rotate the clamping ring in the direction of the arrow until it hits the stop
and hold it there.
▶
Removing the medical device.
▶
Release the clamping ring.
13 / 28