Instructions for use SURGmatic S201 XL Pro - 1.013.7541 SURGmatic S201 L Pro - 1.013.7540
7 Processing steps in accordance with EN ISO 17664 | 7.1 Preparations at the site of use
7 Processing steps in accordance with EN ISO 17664
7.1 Preparations at the site of use
WARNING
Hazard from contaminated products.
Contaminated products are associated with an infection risk.
▶
Take suitable personal protective measures.
▶
Remove the bur from the medical device.
▶
Remove the spray clip for internal cooling.
▶
Remove all residual cement, composite or blood immediately.
▶
Process the medical device as soon as possible after treatment.
▶
The medical device must be dry when transported to processing.
▶
Do not immerse in solutions or the like.
7.2 Non-fixing preliminary cleaning of the spray clip and
spray tube
The non-fixing preliminary cleaning is a central constituent and must be per-
formed prior to the automated processing.
Accessories required:
▪ Demineralized water 30 °C ± 2 °C (86 °F ± 3.6 °F)
▪ Nozzle needle
▪ Brush, e.g. medium-hard toothbrush
▪ Disposable syringe
Note
Before cleaning, take the hose off the spray clip and the contra-angle hand-
piece.
▶
Check the patency of the spray clip and spray tube and clean them with the
nozzle pin (
Mat. no. 0.410.0931
).
▶
Rinse the spray clip and spray tube with at least 20 ml demineralized water
using a disposable syringe.
▶
If the spray clip and spray tube are not patent after the manual rinsing pro-
cedure, the medical device and/or the spray clip must be replaced.
▶
Brush the spray clip and spray tube under running tap water water for at
least 20 seconds using a medium-hard toothbrush.
For validated internal cleaning of the spray clip and spray tube in the cleaning
and disinfecting device, preliminary non-fixing cleaning is required.
17 / 28