Instructions for use SURGmatic S201 XL Pro - 1.013.7541 SURGmatic S201 L Pro - 1.013.7540
7 Processing steps in accordance with EN ISO 17664 | 7.5 Care products and systems - Servicing
See also:
2
Instructions for use KaVo QUATTROcare PLUS
Servicing the chuck
KaVo recommends servicing the chuck system once a week using the chuck
servicing program integrated in the device.
See also:
2
Instructions for use KaVo QUATTROcare PLUS
Note
Handpieces must be taken off the service couplings before the chuck service
can be started and run.
▶
Close the front door and press the chuck service button for at least three
seconds until the spray canister control LED flashes three times consecu-
tively.
ð
The device is in chuck service mode.
▶
Remove the service coupling chuck from the side hatch of the QUATTRO-
care PLUS and attach it to coupling service station four, on the far right. A
MULTIflex adapter must be mounted there.
▶
Press the handpiece together with the guide bush of the chuck to be ser-
viced against the tip of the service coupling.
▶
Press the button marked with the chuck service symbol.
Note
Close the chuck service mode.
Option 1: Place the dental handpieces in the QUATTROcare PLUS 2124 A,
close the front door and start the service procedure.
Option 2: After three minutes with no service procedure running, the device
automatically switches back to normal service mode.
7.5.3 Servicing with KaVo QUATTROcare
Note
QUATTROcare 2104 / 2104 A is no longer included in the current de-
livery program.
Follow-up product:
▶
QUATTROcare PLUS 2124 A
Note
Head can be serviced alone or attached to a shank.
Servicing and cleaning device with expansion pressure for internal cleaning of
inorganic residues and optimum servicing.
(no validated internal cleaning in accordance with German RKI requirements)
KaVo recommends servicing the product as part of the processing after each
use, i.e. after each cleaning, disinfection, and before each sterilization.
▶
Remove the tool from the medical device.
21 / 28