Instructions for use SURGmatic S201 XL Pro - 1.013.7541 SURGmatic S201 L Pro - 1.013.7540
7 Processing steps in accordance with EN ISO 17664 | 7.4 Automated processing
WARNING
Sharp tool in the medical device.
Injury hazard from sharp and/or pointed tool.
▶
Remove the tool.
NOTICE
Never disinfect the handpiece with chlorine-containing products.
Malfunction and material damage.
▶
Only disinfect in the washer disinfector.
NOTICE
Never process this medical device in an ultrasonic device.
Functional damage and property damage.
▶
Clean it in a washer disinfector only.
Note
The spray clip for internal cooling may be additionally cleaned in the ultra-
sonic cleaner.
For validated internal cleaning of the spray clip and spray tube in the cleaning
and disinfecting device, preliminary non-fixing cleaning is required.
7.4.1 Automated internal and external cleaning and
internal and external disinfection
Note
Prior to cleaning or disinfection in a washer disinfector, attach the head to an
appropriate base.
KaVo recommends washer disinfectors in accordance with ISO 15883-1, which
are operated using alkaline cleaning agents.
▶
For program settings as well as cleansers and disinfectants to be used,
please refer to the Instructions for Use of the washer disinfector.
▶
For the spray clip and spray tube, also use the adapter for external spray
channels.
7.4.2 Automated drying
The drying procedure is usually part of the cleaning program of the washer dis-
infector.
Note
Please comply with the instructions for use of the washer disinfector.
▶
In order to prevent impairment of the KaVo medical device, make sure that
the inside and outside of the device is dry after the end of the cycle.
▶
Immediately after drying, lubricate the KaVo medical device with care
agents from the KaVo care system.
19 / 28