2.1.4
Removing the insert and gear from the elbow sleeve
Tool
Name
Assembly pliers (Mat. no. 0.411.3921)
Wrench (Mat. no. 0.411.1881)
Universal bit holder
(Mat. no. 1.002.4577)
Torque handle (Mat. no. 1.003.1524)
180 Ncm
Clamping sleeve (Mat. no. 0.411.4451)
▶ Take hold of the elbow sleeve ① with the circlip pliers and turn the wrench,
universal bit holder and torque handle anticlockwise to unscrew the threaded
ring ④.
▶ Slide the adapter sleeve into the insert ③ and push the spring pin into place.
▶ Pull the insert ③ out of the elbow sleeve ① and remove the adapter sleeve.
▶ Completely remove the gear ② from the elbow sleeve ①.
Technician's instructions SURGmatic S201 L - 1.009.0470
2 Disassembly | 2.1 Disassembly of the SURGmatic S201 L
5/19