Instructions for use QUATTROcare PLUS 2124 A - 1.008.3805

1 User instructions | 1.3 Transportation and storage

▶ Do not use a damaged product.

▶ Report the damage to the shipping company without delay, definitely no later than

7 days after delivery.

Note

If the recipient fails to comply with any of the above-mentioned obligations, the

damage will be considered to have arisen after delivery

(in accordance with CMR law, Chapter 5, Art. 30).

In Germany

If the packaging is visibly damaged on delivery, please proceed as follows:

▶ Leave the product and packaging in the condition in which you received it.

▶ Do not use the product.

▶ Record the loss or damage on the delivery receipt. The recipient and the repre‐

sentative of the shipping company must sign this delivery receipt.

▶ Report the damage to the shipping company.

▶ Report the damage to KaVo.

▶ Consult with KaVo first, before returning a damaged product.

▶ Send the signed delivery receipt to KaVo.

If the product is damaged but there was no discernable damage to the packaging
upon delivery, proceed as follows:

▶ Leave the product and packaging in the condition in which you received it.

▶ Do not use a damaged product.

▶ Report the damage to the shipping company without delay, definitely no later than

7 days after delivery.

▶ Report the damage to KaVo.

Note

If the recipient fails to comply with any of the above-mentioned obligations, the

damage will be considered to have arisen after delivery

(in accordance with the General German Freight Forwarders´ Terms and Conditions, Art. 28)

.

1.3.3 Storage

Note

Please keep the packaging in case you need to return the product for servicing or

repair.

The symbols printed on the outside are for transportation and storage, and have the
following meaning:

Temperature range -20 to 50℃ (-4 to 122

o

F)

 

 

 

 

7 / 36

Summary of Contents for QUATTROcare PLUS 2124 A

Page 1: ...Instructions for use QUATTROcare PLUS 2124 A 1 008 3805 Always be on the safe side...

Page 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Phone 49 0 7351 56 0 Fax 49 0 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com...

Page 3: ...2 3 2 Product specific 12 3 Product description 13 3 1 QUATTROcare PLUS 2124 A 13 3 2 Rating plate 14 3 3 Technical data 14 3 4 Scope of delivery 15 4 First use 17 4 1 Select location 17 4 2 Connect...

Page 4: ...6 2 Safety checks STK 30 7 Troubleshooting 31 8 Tools and consumables 32 9 Terms and conditions of warranty 34 Instructions for use QUATTROcare PLUS 2124 A 1 008 3805 Table of contents 4 36...

Page 5: ...tions AI Assembly instructions TI Technician s instructions SC Safety checks IEC International Electrotechnical Commission RI Repair instructions RK Retrofitting kit AS Assembly set EP Enclosed parts...

Page 6: ...nt or recycling companies Com ply with the comprehensive return system KaVo has had its sales packaging licensed for this purpose Please comply with the regional public waste disposal system 1 3 2 Dam...

Page 7: ...the shipping company Report the damage to KaVo Consult with KaVo first before returning a damaged product Send the signed delivery receipt to KaVo If the product is damaged but there was no discernab...

Page 8: ...008 3805 1 User instructions 1 3 Transportation and storage 8 36 Humidity 5 to 95 non condensing Air pressure 700 to 1 060 hPa Transport upright Protect from impact Protect from moisture 15 kg max Per...

Page 9: ...hazard levels are used in this docu ment to prevent personal and property damage CAUTION CAUTION indicates a hazardous situation that can cause damage to property or mild to moder ate injuries WARNIN...

Page 10: ...egal regulations must be observed during the use of the de vice in particular the following Applicable regulations governing the connection and start up of medical devices Current occupational safety...

Page 11: ...unit and is spe cifically designed to fit this unit only The lubricant spray can is unable to be used manually Note Compliance with the preceding and subsequent re processing steps according to the re...

Page 12: ...e device from heating up to more than 50 122o F CAUTION Contaminated compressed air Damage to and malfunction of the instrument In general ensure dry clean oil free compressed air in accordance with E...

Page 13: ...ATTROcare PLUS 2124 A 3x INTRAmatic service couplings 1x MULTIflex service couplings Control panel and display Door of spray can chamber Chuck service adaptor QUATTROcare Plus Spray canister Collectin...

Page 14: ...ATTROcare PLUS 2124 A SN Year of manufacture serial number REF Material number Fuse protected with T1000 mA Comply with the Instructions for use CE mark CSA mark Labelling according to 2002 96 EC 3 3...

Page 15: ...ed in accordance with EN ISO 7494 2 Operating environment Temperature 15 to 40 o C 59 to 104 o F Humidity 25 to 90 Degree of soiling 2 Permissible to a maximum 2000 m above sea level 6560 feet alti tu...

Page 16: ...of delivery 16 36 Chuck service adaptor QUATTROcare Plus Spray canister Compressed air coupling Instructions for use O ring MULTIflex 4 units O ring INTRAmatic 3 units Compressed air hose Power cable...

Page 17: ...e the unit right next to a source of heat Keep away from ignition sources and do not use open fire close by Ensure sufficient ventilation of the premises CAUTION Danger of overheating Damage to and ma...

Page 18: ...installation location Do not expose to di rect sunlight Do not set up the unit in cabinets Maintain a minimum distance of 50 cm 20 inches from electrical plugs light switches and sources of ignition...

Page 19: ...Spray canister WARNING Use of servicing products other than original KaVo products The use of third party products can shorten the product life cause explosion damage or malfunction Always service KaV...

Page 20: ...sheet of the spray can Note The safety data sheet of the QUATTROcare Plus Spray can be viewed at www kavo com Open the side door of the unit Swing the adaptor on the retention clip outward Place the...

Page 21: ...lly and by touch for absence of leakage The spray can must rest flat against the adaptor There must be no gap between the spray can and the adaptor The connection must be tightly sealed Note If any oi...

Page 22: ...t of the device Place the spray can on the retention clip and unscrew it counterclockwise from the adaptor Close the side door of the unit 4 3 3 Dispose of the spray canister Dispose of the spray can...

Page 23: ...itch on off LED display Start button for chuck service Start button for service procedure 5 2 Open the front flap Note Objects on or in the front door Placing objects on or in the front door can damag...

Page 24: ...ced because their necks are too thin These heads must be serviced together with the corresponding shanks Mount turbines on the MULTIflex service coupling service coupling on the far right and lock the...

Page 25: ...he treatment Note The LED of the spray canister control lighting up indicates that the spray can is empty When the main switch is on the LED lights up Press the start key The automatic servicing proce...

Page 26: ...in the axial direction of the service coupling and take it off Press on the latch with a thumb and take the straight or contra angle handpiece off by simultaneously pulling in the direction of the axi...

Page 27: ...nd fasten it turning the union nut coun terclockwise against the direction of the arrow 5 7 Chuck Servicing Note Handpieces must be taken off the service couplings before the chuck service can be star...

Page 28: ...e chuck to be serviced leading against the tip of the chuck service coupling Note If this is done with insufficient pressure there is a risk of the chuck service adaptor being pushed off the service c...

Page 29: ...ntra angle handpieces and turbines can be changed Hold down the start key Turn the main switch on LEDs 1 to 4 of the LED display light up one after the other The blow out time is as follows for LED 1...

Page 30: ...f the unit using a soft cloth and mild liquid cleanser Remove and clean the collecting pan The collecting pan can be washed in the dishwasher and in the thermodisinfector Temperature max 95 C 203 F Ti...

Page 31: ...is ac tivated Insert original KaVospray canister Notify the KaVo Service Team Flashing LED light is on Bar code is not detected or cannot be read Insert original KaVospray canister Notify the KaVo Ser...

Page 32: ...005 1707 Service coupling BORDEN 2 hole 1 002 1217 O ring INTRAmatic 0 200 6089 O ring MULTIflex 0 200 6098 O ring MULTIflex 0 200 6177 Seal for service coupling for INTRAmatic MULTIflex or heads 0 55...

Page 33: ...8858 Service coupling MORITA Turbines 1 000 6063 Service coupling BIEN AIR 1 001 7860 Service coupling W H 1 001 7888 Presentation Material summary Mat No QUATTROcare Plus Spray canister 2140 P pack...

Page 34: ...ith respect to compensation are excluded In the event of default and gross negligence or intent this shall only ap ply in the absence of mandatory legal regulations to the contrary KaVo cannot be held...

Page 35: ......

Page 36: ...1 008 6955 kb 20150122 13 en...

Reviews: