5.3 Mounting instruments
Note
Remove the handpieces from the unit as soon as the servicing is completed.
Note
The LED display shows the following:
Adaptor on the far left corresponds to LED display 1.
Adaptor on the far right corresponds to LED display 4.
Note
Not every adaptor has to be used.
Note
Heads of the "22" series cannot be serviced because their necks are too thin.
These heads must be serviced together with the corresponding shanks.
▶ Mount turbines on the MULTIflex service coupling (service coupling on the far
right ④) and lock them into place.
▶ Mount straight and contra-angle handpieces on the INTRAmatic service coupling
(service coupling on the far left ①, second from the left ② or third from the left ③)
and lock them into place. Working with LUX handpieces, make sure the latch is in
the correct position.
▶ Mount heads on the service coupling for INTRA heads (for Special accessories,
see chapter on "Consumables and tools"). Push back the locking sleeve ① and
snap it into place.