Instructions for use QUATTROcare PLUS 2124 A - 1.008.3805
4 First use | 4.2 Connect the QUATTROcare PLUS 2124 A
18 / 36
Note
The unit must be set up on a flat, level surface that is insensitive to oil and water.
The unit must not be set up or operated on an angled surface.
Installation requirement:
▪ It is mandatory to maintain a minimal distance of 20 cm (8 inches) ③ between the
unit and wall or cabinet surfaces.
▪ Well-ventilated rooms ① are a suitable installation location. Do not expose to di‐
rect sunlight. Do not set up the unit in cabinets.
▪ Maintain a minimum distance of 50 cm (20 inches) ② from electrical plugs, light
switches, and sources of ignition.
▪ Only operate the device in a clean environment such as that of a laboratory or
dental practice.
4.2 Connect the QUATTROcare PLUS 2124 A
CAUTION
Contaminated compressed air.
Damage to and malfunction of the device.
▶ The compressed air must be free of oil, dirt, and contamination.
▶ If needed: Use a compressor with a dry air system, if required, possible install an
upstream air filter (on the compressor) or blow-out the lines before connecting
them.