Instructions for use QUATTROcare PLUS 2124 A - 1.008.3805
4 First use | 4.1 Select location
4 First use
4.1 Select location
The QUATTROcare Plus Spray canister contains explosive propellant gases. The fol‐
lowing instructions and setup conditions therefore need to be observed when setting
up the unit.
WARNING
Propellant gases of the QUATTROcare Plus spray canister.
Explosive, flammable vapour/air mixtures may be produced during use. Risk of fire
from overheating.
▶ Do not expose the unit to direct sunlight.
▶ Do not place the unit right next to a source of heat.
▶ Keep away from ignition sources, and do not use open fire close by.
▶ Ensure sufficient ventilation of the premises.
CAUTION
Danger of overheating.
Damage to and malfunction of the device.
▶ Do not cover or obstruct the vents and openings on the unit.
▶ Maintain a minimum distance of 20 cm (8 inches) between the unit and the surfa‐
ces of walls or cabinets.
Note
Tests have demonstrated that the unit meets the requirements of EC Directive
2004/108/EC concerning electromagnetic compatibility. Even though the unit does
not emit interfering radiation, other equipment (e.g., ultrasonic cleaners) can inter‐
fere with the unit. Keep the unit away from possible sources of interference.
17 / 36