Instructions for use EXPERTsurg LUX REF 1.008.3500
12 Details on electromagnetic compatibility | 12.3 Guidelines and manufacturer's declaration - electromagnetic resistance to
jamming
Interference immunity
tests
IEC 60601 test levels
Compliance level
Electromagnetic environment -
Guidelines
Wire-based HF interfer-
ence according to IEC
61000-4-6
Wireless HF interfer-
ence according to IEC
61000-4-3
3 V
eff
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 V
eff
3 V/m
Handheld and mobile wireless
devices should not be used at a
shorter distance from the EX-
PERTsurg LUX including cables
than the recommended safe clear-
ance calculated using the appropri-
ate equation for the emission fre-
quency.
Recommended safe distance:
d = [3.5/3]
P
= 1.17
P
d= [3.5/3]
P
=1.17
P
for 80 MHz to
800 MHz
d= [7.0/3]
P
= 2.33
P
for 800 MHz
to 2.5 GHz
where P is the maximal nominal
power of the transmitter in watts
(W) as specified by the transmitter
manufacturer and d is the recom-
mended safe clearance in metres
(m).
The field strength of stationary wire-
less radio transmitters as measured
locally
a
should be lower than the
conformance level at all frequen-
cies.
b
Interference is possible in the vicin-
ity of devices that bear the following
symbol.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not be applicable in every case. The propagation of
electromagnetic waves is subject to absorption and reflection by buildings, objects,
and people.
a
The field strength of stationary transmitters such as base stations of mobile tele-
phones and land mobile radio devices, amateur radio stations, AM and FM radio and
television broadcasting stations cannot be determined based on theoretical considera-
tions. A site study should be considered to determine the electromagnetic environment
in terms of stationary transmitters. If the measured field strength at the site, at which
the EXPERTsurg LUX is used, exceeds the compliance levels shown above, the EX-
PERTsurg LUX should be monitored to demonstrate proper function. If any uncom-
mon performance characteristics are observed, additional measures may be required,
such as, e.g., changing the orientation or using a different location for the EX-
PERTsurg LUX.
b
In the frequency range of 150 kHz to 80 MHz, the field strength should be less than
3V
eff
V/m.
69 / 72