Instructions for use EXPERTsurg LUX REF 1.008.3500
5 Operation | 5.8 Foot control
▶ Press the plus or minus keys to start the one-touch calibration.
ð
The display shows "Press foot control".
CAUTION
The motor starts at full speed.
Risk of injury.
▶ Hold the motor firmly or put it in a safe holder during the calibration.
▶ Press the foot control and hold it down until the display shows that the calibration
has been successful by displaying the message, "Measurement done".
▶ If you release the foot control before the display shows that the calibration was
successful, press the foot control again until the display shows that the calibration
was successful.
▶ Press the back key to terminate the calibration and to return to the selection of
device settings.
If an unsuitable or defective handpiece was used in the calibration, the calibration is
discontinued and the error message, "Measurement failed - Non-permissible current",
is shown.
▶ Press the back key to terminate the failed calibration.
See also:
2
8 Troubleshooting, Page 61
5.8 Foot control
5.8.1 Changing the speed, coolant flow, and direction of motor rotation
▶ Press the speed key with your foot to start the motor and increase the speed.
49 / 72