Instructions for use EXPERTsurg LUX REF 1.008.3500
12 Details on electromagnetic compatibility | 12.4 Recommended safe distance between portable and mobile HF telecom-
munications equipment and the EXPERTsurg LUX
70 / 72
12.4 Recommended safe distance between portable and mobile HF
telecommunications equipment and the EXPERTsurg LUX
The EXPERTsurg LUX is for use in an electromagnetic environment like the one cited
below. The customer or user of the EXPERTsurg LUX can help to avoid electromag-
netic faults by keeping to the minimum distance apart between portable and mobile
HF-telecommunication devices (transmitters) and the EXPERTsurg LUX - dependent
on the output lines for the communication device - as given below.
Rated power of the trans-
mitter in W
150 kHz to 80 MHz
d=1.17
P
80 MHz to 800 MHz
d=1.17
P
800 MHz to 2.5 GHz
d=2.33
P
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.17
1.17
2.33
10
3.70
3.70
7.37
100
11.70
11.70
23.30
For transmitters whose maximum rated power is not included in the above table, the
recommended safe distance d in metres (m) can be calculated using the equation for
the respective column, where P is the maximum rated power of the transmitter in
Watts (W) as specified by the manufacturer.
Note 2: These guidelines may not be applicable in every case. The spread of electro-
magnetic waves is absorbed and reflected by buildings, objects and people.
Comment 1: To calculate the recommended safe distance from transmitters with a fre-
quency range of 80 MHz to 2.5 GHz, an additional factor of 10/3 was used to reduce
the probability that a mobile/portable communication unit that is inadvertently brought
into the patient area would cause malfunction.