Instructions for use EXPERTsurg LUX REF 1.008.3500
5 Operation | 5.4 Setting and executing program steps
5.4.1 Factory settings
The following program steps are pre-set at the factory:
Programme
step
Icon
Activity
Speed [rpm]
Torque [Ncm]
Transmission
ratio
Coolant flow
1
Marking
200 – 2,000
500 (D)
5 – 20
10 (D)
16:1
27:1
20:1 (D)
0 – 4
2 (D)
2
Pilot drilling
200 – 2,000
500 (D)
5 – 20
10 (D)
16:1
27:1
20:1 (D)
0 – 4
2 (D)
3
Template
drilling
200 – 2,000
500 (D)
5 – 20
10 (D)
16:1
27:1
20:1 (D)
0 – 4
2 (D)
4
Tapping
15 – 50
20 (D)
5 – 80
25 (D)
16:1
27:1
20:1 (D)
0 – 4
2 (D)
5
Placing im-
plant
15 – 50
20 (D)
5 – 80
25 (D)
16:1
27:1
20:1 (D)
0 – 4
0 (D)
6
Setting a clos-
ure cap
15 – 50
20 (D)
5 – 15
8 (D)
16:1
27:1
20:1 (D)
0 – 4
0 (D)
7
Free use
300 – 40,000
40,000 (D)
0.15 – 5.5
3 (D)
1:1
0 – 4
2 (D)
20 – 2,000
5 – 80
16:1
15 – 2,000
5 – 80
20:1
15 – 1,200
5 – 80
27:1
8
Rinsing func-
tion
–
–
–
–
39 / 72
Summary of Contents for EXPERTsurg LUX 1.008.3500
Page 1: ...Instructions for use EXPERTsurg LUX REF 1 008 3500...
Page 71: ......