Instructions for use EXPERTsurg LUX REF 1.008.3500
4 First use | 4.5 Connecting the coolant container and hose set
① Coolant hose
② Clip
③ Hose fixation
④ Lock
⑥ Insertion needle
⑦ Hose clamp
▶ Close the hose clamp ⑦ of the hose set.
▶ Attach the coolant hose ① to the straight or contra-angle handpiece.
▶ Place the coolant hose ① tightly, without loops or kinks, against the outside of the
motor cable and attach it in regular intervals using the enclosed clips ②.
▶ Plug the hose fixation ③ into the opening of the device with the blue border until
the hose fixation ③ snaps in.
Note
The unit recognises the hose fixation. If the unit fails to recognise the hose fixation
or if it is not plugged in, the coolant supply symbol is shown with a yellow back-
ground.
25 / 72
Summary of Contents for EXPERTsurg LUX 1.008.3500
Page 1: ...Instructions for use EXPERTsurg LUX REF 1 008 3500...
Page 71: ......