Instructions for use DIAGNOcam 2170 U
9 Information concerning the electromagnetic compatibility according to EN IEC 60601-1-2 | 9.3 Immunity to elec-
tromagnetic interference
Interference im-
munity tests
EN 60601 test level
Compliance level
Electromagnetic en-
vironment -
Guidelines
Electrostatic discharge
(ESD) according to EN
61000-4-2
± 6 kV contact discharge
± 8 kV atmospheric dis-
charge
± 2/4/6 kV contact dis-
charge
± 2/4/8 kV atmospheric
discharge
Floors should be made of
wood or concrete or be
fitted with ceramic tiles.
If the floor is fitted with
synthetic material, the
relative humidity must
be at least 30%.
Fast transient electrical
interference / bursts ac-
cording to EN 61000-4-4
± 2 kV for power lines
± 1 kV for input and out-
put lines
± 2 kV for power lines
The quality of the supply
voltage should corres-
pond to that of a typical
business or
hospital environment.
Surges according to EN
61000-4-5
± 1 kV push-pull voltage
± 2 kV common mode
voltage
± 1 kV push-pull voltage
± 2 kV common mode
voltage
The quality of the supply
voltage should corres-
pond to that of a typical
business or
hospital environment.
Voltage interruptions,
short-term interruptions
and fluctuations of the
supply voltage according
to EN 61000-4-11
< 5% U
T
(> 95% interruption)
for ½ period
40% U
T
(60% interruption)
for 5 periods
70% U
T
(30% interruption)
for 25 periods
< 5% U
T
(> 95% interruption)
for 5 s
(250 periods)
< 5% U
T
(> 95% interruption)
for ½ period
40% U
T
(60% interruption)
for 5 periods
70% U
T
(30% interruption)
for 25 periods
< 5% U
T
(> 95% interruption)
for 5 s
(250 periods)
The quality of the supply
voltage should corres-
pond to that of a typical
business or hospital en-
vironment. If the user
needs the DIAGNOcam
2170 U to work even if
the power supply is in-
terrupted, we recom-
mend supplying energy
to the DIAGNOcam 2170
U from an uninterrupt-
ible power supply or bat-
tery.
Magnetic field at a sup-
ply frequency (50/60 Hz)
according to EN
61000-4-8
3 A/m
3 A/m
Magnetic fields at the
mains frequency should
correspond to typical
values in a business and
hospital environment.
Note: U
T
is the alternating mains voltage before the test level is used.
9.3 Immunity to electromagnetic interference
The DIAGNOcam 2170 U is designed for use in an environment of the kind spe-
cified below. The customer or the user of the DIAGNOcam 2170 U should en-
sure that the device is used in an environment of the specified type.
33 / 38
Summary of Contents for DIAGNOcam 2170 U
Page 1: ...Instructions for use DIAGNOcam 2170 U...
Page 36: ......
Page 37: ......
Page 38: ...1 011 5160 bd 20220323 03 en...