Instructions for use DIAGNOcam 2170 U
6 Reprocessing methods according to ISO 17664 | 6.7 Control and functional checks
CAUTION
Moisture
Non-sterility
▶
Ensure dryness. Autoclaves with a after-vacuum ensure dryness! In addi-
tion, drying can be accelerated through a 10 minute drying phase with the
autoclave door open.
CAUTION
Product damage due to inappropriate cleaning
Do not place the tip in an ultrasonic bath.
Only the tips can be sterilised.
Sterilisation should directly follow cleaning and disinfection.
Bag the tips prior to sterilisation.
The KaVo products released for sterilisation have a maximum temperature res-
istance of 138
℃
.
▶
Sterilising the probes in an autoclave:
Select a suitable procedure (depending on the available autoclave) from the fol-
lowing sterilisation processes:
▪ Autoclave with pre-vacuum:
- at least 3 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4 °F)
Drying time: 10 min.
▪ Autoclave using the gravity method:
- at least 10 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4 °F) or
alternatively
Drying time: 10 min.
Store the tips in a bag
Note
When sterilising several handpieces in a single sterilisation cycle, do not ex-
ceed the maximal load of the steriliser.
6.7 Control and functional checks
6.7.1 General
▶
Check for cleanliness.
6.7.2 Checking the tips
▶
Hold tips against a light source (e.g. daylight source) and check the optical
fibres seen in the sleeve for contamination and defects.
29 / 38