8
Troubleshooting
Malfunction
Cause
Remedy
The suction is too low (also in the
highest level of the suction control).
The filter elements are dirty.
▶ Clean or replace the filter ele‐
ments
The exhaust paths or air exit.
▶ Collect the exhaust paths (espe‐
cially air louvers when installed)
and eliminate hindrances.
The suction fan does not work even
though the power switch is on.
Fan switch is in postion "0" (off).
▶ Push the fan switch into position
"1" (manual mode) or down to
position "2" (automatic mode).
The fan switch is in position "2" (au‐
tomatic mode), and the set turn-on
time has not been reached.
▶ Move the fan switch up to posi‐
tion "1" (manual operation), or
change the turn on time of auto‐
matic mode.
The fuse is blown for the suction fan. ▶ Only a KaVo technician or
trained employee may exhange
the fuse.
Follow the safety instructions.
Instructions for use Type 5650, 5651, 5652, 5653
8 Troubleshooting
25/29
Summary of Contents for 5650
Page 1: ...Documents Type 5650 5651 5652 5653 Always on the safe side...
Page 3: ...Gebrauchsanweisung Type 5650 5651 5652 5653 Immer auf der sicheren Seite...
Page 30: ......
Page 31: ...0 488 5434 kf 20090511 02 de...
Page 32: ...Instructions for use Type 5650 5651 5652 5653 Always be on the safe side...
Page 59: ......
Page 60: ...0 488 5434 kf 20090511 02 en...
Page 61: ...Mode d emploi Type 5650 5651 5652 5653 En toute s curit...
Page 88: ......
Page 89: ...0 488 5434 kf 20090511 02 fr...
Page 90: ...Instrucciones de uso Type 5650 5651 5652 5653 Siempre a lo seguro...
Page 117: ......
Page 118: ...0 488 5434 kf 20090511 02 es...
Page 119: ...Istruzioni per l uso Type 5650 5651 5652 5653 Sempre dalla parte della sicurezza...
Page 146: ......
Page 147: ...0 488 5434 kf 20090511 02 it...
Page 148: ......