4
Fitting and connecting
4.1
Check the location
The necessary minimum room height is:
▪ 2500 mm with a direct vertical connection (A)
▪ 2750 mm with a flanged-on elbow (B)
The pipe diameter for the exhaust must be at least 200 mm.
Note
The overall length of the exhaust pipe should be kept short to ensure ef‐
fectiveness. Consequently, avoid bends and narrowing diamters of the
exhaust pipes.
Note
The provided screw sets are only suitable for walls made of steel-rein‐
forced concrete, building tile and non-perforated limestone.
With walls made of the following materials, use injection anchors of type "F/M 10N"
(Mat. no. 0.224.6666) and hex head screws "M10 x 20" according to DIN 933
(Mat. no. 0.251.0202):
▪ Cored brick
▪ Perforated limestone
▪ Full brick of light concrete (pumice)
▪ Hollow block of light concrete
▪ Aerated concrete
▶ Check that the rated voltage frequency match the specifications on the rating
plate.
▶ Note the minimum room height.
▶ Ensure sufficient ventilation.
Instructions for use Type 5650, 5651, 5652, 5653
4 Fitting and connecting | 4.1 Check the location
15/29
Summary of Contents for 5650
Page 1: ...Documents Type 5650 5651 5652 5653 Always on the safe side...
Page 3: ...Gebrauchsanweisung Type 5650 5651 5652 5653 Immer auf der sicheren Seite...
Page 30: ......
Page 31: ...0 488 5434 kf 20090511 02 de...
Page 32: ...Instructions for use Type 5650 5651 5652 5653 Always be on the safe side...
Page 59: ......
Page 60: ...0 488 5434 kf 20090511 02 en...
Page 61: ...Mode d emploi Type 5650 5651 5652 5653 En toute s curit...
Page 88: ......
Page 89: ...0 488 5434 kf 20090511 02 fr...
Page 90: ...Instrucciones de uso Type 5650 5651 5652 5653 Siempre a lo seguro...
Page 117: ......
Page 118: ...0 488 5434 kf 20090511 02 es...
Page 119: ...Istruzioni per l uso Type 5650 5651 5652 5653 Sempre dalla parte della sicurezza...
Page 146: ......
Page 147: ...0 488 5434 kf 20090511 02 it...
Page 148: ......