Instructions for use MASTERtorque LUX M9000 L - 1.008.7900 MASTERtorque LUX M9000 LS - 1.008.5400
5 Operation | 5.3 Removing the medical device
18 / 34
▶
Pull on it to make sure that the medical device is securely affixed to the
coupling.
5.3 Removing the medical device
▶
Grasping the coupling, twist the medical device slightly and pull it off.
5.4 Inserting the dental bur
Note
Only use carbide cutters or diamond grinders that correspond to DIN EN ISO
1797 type 3, are made of steel or hard metal and meet the following criteria:
- Shaft diameter: 1.59 to 1.60 mm (0.0626 in to 0.0629 in)
- Overall length: max. 25 mm
- Shaft clamping length: at least 11 mm
- Blade diameter: max. 2 mm (0.0787 in)
WARNING
Use of non-approved dental burs.
Injury to the patient or damage to the medical device.
▶
Comply with the instructions for use and the intended use of the dental bur.
▶
Only use dental burs that do not deviate from the specified data.
CAUTION
Dental bur with worn or damaged shafts.
Risk of injury, dental bur may fall out during treatment.
▶
Never use a dental bur with damaged or worn shafts.
CAUTION
Contaminated, sharp-edged dental bur.
Infections or cuts.
▶
Use gloves or a finger guard when you test, insert and remove the dental
bur.
CAUTION
Defective clamping system.
Risk of injury, dental bur may fall out during treatment.
▶
Pull on the dental bur to check if the clamping system works properly and if
the dental bur is firmly clamped.
NOTICE
Dental bur with worn or damaged shafts.
Material damage to the chuck system, dental bur is difficult or impossible to re-
move from the chuck system.
▶
Never use a dental bur with damaged or worn shafts.