7 Reprocessing steps in accordance with ISO 17664 | 7.2 Manual Reprocessing
▶
Hold the can vertically.
▶
Press the spray button three times for 2 seconds each time.
▶
Remove the medical device from the spray attachment and let the cleanser
act for 1 minute.
See also:
2
KaVo CLEANspray Instructions for Use
▶
If a manual external and internal disinfection do not follow directly, dry the
medical device with KaVo DRYspray.
See also:
2
7.2.3 Manual external disinfection
WARNING
Incomplete disinfection.
Infection hazard.
▶
Only use disinfection procedures that are verified to be bactericidal, fungi-
cidal and virucidal.
▶
If the disinfectants used do not meet these requirements, the process must
be concluded by disinfection of the unit(s) without packaging using a steam
steriliser.
NOTICE
Never disinfect the medical device with chloride-containing products.
Malfunction and material damage.
▶
Only disinfect in a washer disinfector or manually.
KaVo recommends the following products based on compatibility of the materi-
als. The microbiological efficacy must be ensured by the disinfectant manufac-
turer and proven by an expert opinion.
Approved disinfectants:
▪ CaviWipes and CaviCide made by Metrex (intermediate disinfection)
▪ Mikrozid AF made by Schülke & Mayr (liquid or cloths)
▪ FD 322 made by Dürr
Consumables required:
▪ Cloths for wiping the medical device.
▶
Spray the disinfectant on a cloth, then wipe down the medical device and
allow the disinfectant to act according to the instructions of the
disinfectant manufacturer.
▶
Follow the instructions for use of the disinfectant.