Instructions for use MASTERtorque LUX M9000 L - 1.008.7900 MASTERtorque LUX M9000 LS - 1.008.5400
6 Troubleshooting | 6.4 Changing the water filter
CAUTION
Hazard from insufficient spray water.
Overheating of the medical device and damage to the tooth.
▶
Check the spray water channels and clean the spray nozzles with the nozzle
needle
Mat. no. 0.410.0921
if necessary.
▶
Check the water filter and exchange if necessary.
6.4 Changing the water filter
WARNING
Hazard from contaminated products.
Infection hazard for care provider and patient.
▶
Prior to initial commissioning and after each use, reprocess the product and
accessories.
CAUTION
Overheating of the tooth due to insufficient amount of cooling water.
Insufficient spray water can cause the medical device to overheat and damage
the pulp and tooth.
▶
Check the water filter and replace it, if needed.
▶
Check spray water channels and if necessary clean spray nozzles with the
nozzle needle (
Mat. no. 0.410.0921
).
▶
Unscrew the sleeve
①
in counterclockwise direction from the insert
②
and
pull it off.
▶
Unscrew the water filter
③
with the wrench (Mat. no. 1.002.0321) and take
it out
▶
Insert the new filter (Mat. no. 1.002.0271) and screw it in with the wrench
▶
Place the sleeve
①
on the insert
②
, and screw it tight in clockwise direction.
21 / 34