A higher drive pressure will be reduced automatically by the medical device.

4.4

   

Check O-rings

CAUTION

Missing or damaged O-rings.

Malfunctions and premature failure.

▶ Make sure that all O-rings are on the coupling and undamaged.

Number of available O-rings: 5

 

 

 

Instructions for use GENTLEsilence LUX 8000B - 1.000.0800 | GENTLEsilence 8000C - 1.000.2800
4 First use | 4.4 Check O-rings

9/18

Summary of Contents for 1.000.0800

Page 1: ...Instructions for use GENTLEsilence LUX 8000B 1 000 0800 GENTLEsilence 8000C 1 000 2800 Always be on the safe side ...

Page 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Phone 49 0 7351 56 0 Fax 49 0 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com ...

Page 3: ...7664 11 6 1 Preparations at the site of use 11 6 2 Cleaning 11 6 2 1 Manual cleaning external 11 6 2 2 Automated external cleaning 11 6 2 3 Manual cleaning internal 12 6 2 4 Automated internal cleaning 12 6 3 Disinfection 12 6 3 1 Manual disinfection external 13 6 3 2 Manual disinfection internal 13 6 3 3 Machine disinfection external and internal 13 6 4 Drying 14 6 5 Care products and systems Ser...

Page 4: ...or users and service technicians Can be steam sterilised at 134 C 1 C 4 C 273 F 1 6 F 7 4 F Thermodisinfectable CE mark European Community A product bearing this mark meets the requirements of the applicable EC directive Action request Target group This document is intended for dentists and their assistants The section on start ing up is also intended for service technicians Instructions for use G...

Page 5: ...ous situation that can cause damage to property or mild to moderate injuries WARNING WARNING indicates a hazardous situation that can lead to serious or fatal injury DANGER DANGER indicates a maximal hazard due to a situation that can directly cause death or fatal injury 2 2 Safety instructions WARNING Hazards for the care provider and the patient In the case of damage irregular running noise exce...

Page 6: ...dication If the soft tissue in the oral cavity is injured the compressed air may enable sep tic substances to enter the tissue Treatment must be discontinued with instruments operated by compressed air when soft tissue is damaged in the oral cavity CAUTION Burning hazard from hot instrument cover If the instrument overheats burns may arise in the oral area Never contact soft tissue with the instru...

Page 7: ...products and care systems described therein must be used KaVo recommends specifying a service interval at the dental office for a licensed shop to clean serv ice and check the functioning of the medical device This service interval depends on the frequency of use and should be adjusted accordingly Service may only be carried out by KaVo trained repair shops using original KaVo replacement parts In...

Page 8: ...nt national statutory regulations Proper use According to these regulations this medical device may only be used for the de scribed application by a knowledgeable user The following must be observed the applicable health and safety regulations the applicable accident prevention regulations these instructions for use According to these regulations it is the responsibility of the user to only use eq...

Page 9: ...ous to start up the medical device after it has been stored strongly refrigerated This can cause the medical device to malfunction Prior to start up very cold products must be heated to a temperature of 20 C to 25 C 68 F to 77 F Temperature 20 C to 70 C 4 F to 158 F Relative humidity 5 RH to 95 RH absence of condensation Air pressure 700 hPa to 1060 hPa 10 psi to 15 psi Protect from moisture Instr...

Page 10: ... spray ring on the MULTIflex LUX MULTIflex LED coupling 4 2 Check the amount of water CAUTION Overheating of the tooth due to lack of cooling water Thermal damage to the dental pulp Adjust the water amount for the spray cooling to a minimum of 50 cm3 min 4 3 Check the pressure CAUTION Compressed air connection on devices Dirty and moist compressed air causes premature bearing wear Ensure that the ...

Page 11: ... rings CAUTION Missing or damaged O rings Malfunctions and premature failure Make sure that all O rings are on the coupling and undamaged Number of available O rings 5 Instructions for use GENTLEsilence LUX 8000B 1 000 0800 GENTLEsilence 8000C 1 000 2800 4 First use 4 4 Check O rings 9 18 ...

Page 12: ...sufficient spray water can cause the medical device to overheat and damage the tooth Check spray water channels and if necessary clean spray nozzles with the nozzle needle Mat no 0 410 0921 Clean the water passage in the spray nozzles by using the nozzle needle Mat no 0 410 0921 5 3 Change the water filter Unscrew the sleeve anticlockwise from the insert and pull off Pull back the insert slightly ...

Page 13: ...quired accessories Tap water 30 C 5 C 86 F 10 F Brush such as a medium hard toothbrush Brush off under flowing tap water 6 2 2 Automated external cleaning KaVo recommends thermodisinfectors in accordance with EN ISO 15883 1 which are operated with alkaline cleaning agents with a pH value of max 10 e g Miele G 7781 G 7881 Validation was carried out with Programme VARIO TD cleaning agent neodisher m...

Page 14: ...sinfec tors in accordance with EN ISO 15883 1 6 2 4 Automated internal cleaning KaVo recommends thermodisinfectors in accordance with EN ISO 15883 1 which are operated with alkaline cleaning agents with a pH value of max 10 e g Miele G 7781 G 7881 Validation was carried out with Programme VARIO TD cleaning agent neodisher mediclean neutralisation agent neodisher Z and rinsing agent neodisher mielc...

Page 15: ...e agents from the KaVo care system Follow the instructions for use of the disinfectant 6 3 3 Machine disinfection external and internal KaVo recommends thermodisinfectors in accordance with EN ISO 15883 1 which are operated with alkaline cleaning agents with a pH value of max 10 e g Miele G 7781 G 7881 Validation was carried out with Programme VARIO TD cleaning agent neodisher mediclean neutralisa...

Page 16: ...and malfunctions from improper servicing and care Reduced product life Perform proper care regularly Note KaVo only guarantees that its products will function properly when the care prod ucts used are those listed as accessories as they were tested for proper use on our products 6 5 1 Servicing with KaVo Spray KaVo recommends servicing the project twice daily at noon and in the evening after hours...

Page 17: ... and care unit with expansion pressure for effective cleaning and care KaVo recommends servicing the product as part of the reprocessing after each use i e after each cleaning disinfection and before each sterilisation Remove the cutter or grinder Servicing the product KaVo QUATTROcare plus spray can KaVo recommends cleaning and servicing the chuck system once a week See the KaVo QUATTROcare instr...

Page 18: ...uct Immediately remove the product from the steam steriliser after the sterilisa tion cycle The KaVo medical device has a maximum temperature resistance up to 138 280 4 F Select a suitable procedure depending on the available autoclave from the fol lowing sterilisation processes Autoclave with three times pre vacuum at least 3 minutes at 134 C 1 C 4 C 273 F 1 6 F 7 4 F Autoclave using the gravity ...

Page 19: ...Insert for turbines 0 411 9902 Cleanpac 10 units 0 411 9691 MULTIflex spray head nozzle 0 411 9921 STERIclave bags 0 411 9912 Material summary Mat no Adaptor INTRAmatic CLEANspray and DRYspray 1 007 1776 KaVo CLEANspray 2110 P 1 007 0579 KaVo DRYspray 2117 P 1 007 0580 KaVo Spray 2112 A 0 411 9640 ROTAspray 2 2142 A 0 411 7520 QUATTROcare plus Spray 2140 P 1 005 4525 Instructions for use GENTLEsil...

Page 20: ...e non compliance with operating maintenance or connection instructions calcination or corrosion contaminated air or water supplies or chemical or electrical factors deemed abnormal or impermissible in accordance with KaVo s instructions for use or other manufacturer s instructions The warranty granted does not usually ex tend to lamps light conductors made of glass and glass fibres glassware rubbe...

Page 21: ......

Page 22: ...1 003 6890 kb 20101109 02 en ...

Reviews: