6.6
Packaging
Note
The sterilisation bag must be large enough for the handpiece so that the bag is
not stretched.
The quality and use of the sterilisation packaging must satisfy applicable stand‐
ards and be suitable for the sterilisation procedure!
▶ Weld each medical device individually in a sterilised item package!
6.7
Sterilisation
Sterilisation in a steam steriliser (autoclave) in accordance with EN
13060 / ISO 17665-1
CAUTION
Premature wear and malfunctions from improper servicing and care.
Reduced product life.
▶ Before each sterilisation cycle, service the medical device with KaVo care
products.
CAUTION
Contact corrosion due to moisture.
Damage to product.
▶ Immediately remove the product from the steam steriliser after the sterilisa‐
tion cycle!
The KaVo medical device has a maximum temperature resistance up to 138 ℃
(280.4 °F).
Select a suitable procedure (depending on the available autoclave) from the fol‐
lowing sterilisation processes:
▪ Autoclave with three times pre-vacuum:
– at least 3 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4 °F)
▪ Autoclave using the gravity method:
– at least 10 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4 °F)
▶ Use according to the manufacturer's Instructions for Use.
6.8
Storage
▶ Reprocessed products should be stored protected from dust with minimum ex‐
posure to germs in a dry, dark and cool space.
▶ Comply with the expiry date of the sterilised items.