5
Troubleshooting
5.1
Exchanging the O-rings on the coupling on the supply hose
CAUTION
Hazard from improper care of the O-rings.
Malfunctions or complete failure of the medical device.
▶ Do not use Vaseline or other grease or oil.
Note
The O-ring on the coupling may only be lubricated with a cotton ball wet with KA‐
VO spray.
▶ Press the O-ring between your fingers to form a loop.
▶ Shove the O-ring to the front, and remove it.
▶ Insert new O-rings into the grooves.
5.2
Cleaning the spray nozzle.
CAUTION
Hazard from insufficient amount of spray water.
Insufficient spray water can cause the medical device to overheat and damage
the tooth.
▶ Check spray water channels and if necessary clean spray nozzles with the
nozzle needle (Mat. no. 0.410.0921).
▶ Clean the water passage in the spray nozzles by using the nozzle needle
(Mat. no. 0.410.0921).
5.3
Change the water filter
▶ Unscrew the sleeve ① anticlockwise from the insert ② and pull off.
▶ Pull back the insert ② slightly.
▶ Use tweezers to remove the water filter ③ from the insert ②.
▶ Insert a new water filter.
▶ Pull back the insert ② slightly.
▶ Place the sleeve ① on the insert ②, and screw it tight clockwise.
Instructions for use GENTLEsilence LUX 8000B - 1.000.0800 | GENTLEsilence 8000C - 1.000.2800
5 Troubleshooting | 5.1 Exchanging the O-rings on the coupling on the supply hose
10/18