6
Preparation methods according to ISO 17664
6.1
Preparations at the site of use
WARNING
Hazard from nonsterile products.
There is a risk of infection from contaminated medical devices.
▶ Take suitable personal protective measures.
▶ Remove all residual cement, composite or blood without delay.
▶ Recondition the medical device as soon as possible after treatment.
▶ Remove the cutter or grinder from the medical device.
▶ The medical device must be dry when transported for reconditioning.
▶ Do not place it in a solution or similar.
6.2
Cleaning
CAUTION
Malfunctions from cleaning in the ultrasonic unit.
Defects in the product.
▶ Only clean manually or in a thermodisinfector.
6.2.1
Manual cleaning - external
Required accessories:
▪ Tap water 30°C ± 5°C (86°F ± 10°F)
▪ Brush such as a medium hard toothbrush
▶ Brush off under flowing tap water.
6.2.2
Automated external cleaning
KaVo recommends thermodisinfectors in accordance with EN ISO 15883-1, which
are operated with alkaline cleaning agents with a pH value of max. 10 (e.g. Miele
G 7781 / G 7881 – Validation was carried out with Programme "VARIO-TD",
cleaning agent "neodisher
®
mediclean", neutralisation agent "neodisher
®
Z" and
rinsing agent "neodisher
®
mielclear" and only applies to the material compatibility
with KaVo products).
▶ For program settings as well as cleansers and disinfectants to be used, please
refer to the Instructions for Use of the thermodisinfector (complying with max.
pH value of 10).
Instructions for use GENTLEsilence LUX 8000B - 1.000.0800 | GENTLEsilence 8000C - 1.000.2800