3. Display Description and operation Procedure
23
3.4.6 Utility
Select UTILITY on the menu screen to display the screen shown on the
left.
In UTILITY you can perform preparation for a measurement and
configure settings.
Press [
▲
] [
▼
] keys to move the cursor to the item to be set, and then
press [SET] key to configure the setting.
CLEANING
TIME ADJUST: Time Setting
K FACTOR: The Mass Concentration Conversion Coefficient setting
LCD/BEEP:
LCD & Beep Sound setting
ANALOG OUT/ALARM: Analog Output & Alarm setting
TWA MODE: TWA Measurement Mode Setting
To go back to the MENU screen, press [MENU] key.
A)
Cleaning
To clean the instrument
→
Refer to
2.3 Cleaning
.
To start cleaning, put the rubber cap on the inlet. Confirm that the
sensitivity check knob is at the “ZERO” side.
Press [START/STOP] key to start cleaning.
Press [MENU] key to go back to the UTILITY screen.
Perform cleaning for approximately 5 minutes.
*
Required cleaning time is different depending on the use
environment.
*
If an error occurs during calibration and background adjustment after
cleaning for more than 10 minutes, the internal cleaning is required.
Please contact your distributor or KANOMAX service center.
[START/STOP] Key
Push
“
START/STOP
”
Key to stop cleaning. Push MENU Key to go
back to
UTILITY
Menu.