Technical manual
KANMED Operatherm OP3
Page
30
of
34
PAD Disassembling instructions:
NOTE: No repairs are possible on the PAD connector circuit board. It is not possible to replace
the board either since it is factory calibrated with the pad and the connecting cable. Only the
casing can be replaced if damaged.
Remove all screws (4 pcs) from the
bottom plate of the Connector assembly.
Remove the bottom part. Replace the
PAD Connector Casing if required and
refit the screws.
Pls observe to use proper ESD
protection!