Technical manual
KANMED Operatherm OP3
Page
17
of
34
Heating Pad Data Summary
Information about the connected Pad. Pad serial number and
type, SW version, Calibration information, Running hours and
No of starts (operating sessions).
“Menu” (up arrow) returns the system to the Technical data
menu.
NOTE
:
Also available in the Customer Software version.
Heating Pad Calibration Data
Information about the connected Pad.
Factory set Calibration values for each of the 8 elements.
“Menu” (up arrow) returns the system to the Technical data
menu.
NOTE
:
The top value corresponds to the resistance measured for each individual element, at the
calibration temperature displayed as the bottom figure. This function is also available in the
Customer Software version.
System Measurements
This function measures each of the eight channels separately
for deep analyze of the system measurement accuracy and the
condition of each of the elements in the connected heating
pad.
Press the [‹] or [›] buttons to change the channel to study.
[v] (down arrow) changes to a screen where all 8 channels are
shown simultaneously.
“Menu” (up arrow) returns the system to the Technical data
menu.
NOTE
:
This is a function that is intended to analyze a Control unit’s ability to measure correctly on
each of the separate channels. If a Calibration Adapter is connected it shall measure the reference
resistance to a value better than 0,03 Ohm. It can also be used to measure the Heating Pad
resistance on each of the separate heating elements. This function is also available in the Customer
Software version.