KM9106 Operators Manual
6
4. NORMAL START UP SEQUENCE
4.1 Every Time You Use The Analyser
BEFORE SWITCH-ON CHECK THAT:
the Oxygen sensor is connected
the particle filter is not dirty
the sulphur filter is fitted for heavy oil or coal
the water trap and probe line are empty of water
all hose connections, etc, are properly made
the paper roll is fitted
the probe is sampling ambient FRESH air
the water trap is vertical
the flue temperature is connected
the instrument is placed on a clean, flat, level surface
Switch ON the instrument by pressing
4.2 Automatic Calibration
During this sequence the analyser pumps fresh air into the sensors to allow toxic
sensors to be set to zero and the Oxygen sensor to be set to 20.9 %.
After switch-on the analyser will
briefly display the Kane logo and
telephone numbers:-
777
KANE-MAY
777
TEL +44 (0) 1707 375550
FAX +44 (0) 1707 393277
Followed by the logger menu screen :-
9
9
SELECT FUNCTION
9
9
. . LOGGER CONTROL
2 . . QUINTOX CONTROL
Use the and keys to
position the cursor over
2 . .Quintox
Control.
Press to access Quintox
Control