KM9106 Operators Manual
28
CALIBRATE
SENSOR
Select using or key and
YES
NO
Select YES
SET ZERO ?
NO
Parameter
Controls Options
SET ZERO
Select using or key and
YES
NO
Select YES
SELECT SENSOR
CO -------- . . -17
Parameter
Controls Options
SELECT
SENSOR
Select using or key and
LIST OF
FITTED
SENSORS
DISPLAYED
In this example we have selected the CO sensor. The pump will now run if it has been
turned off, this is to draw
fresh air
through the instrument to allow the sensors to be
zeroed.
Ignore the reading next to the sensor
Ensure the instability reads zero
SET ZERO LEVEL
CO -------- . . -17
INSTABILITY 5
7
CALIBRATING
77
Once the instability has reached zero press
to accept the zero calibration
and return to the main display.
If the instability does not reach zero then the instrument will show
FAULT
. Contact
Kane International or Authorised Distributor for advice.