KM9106 Operators Manual
7
And show the countdown screen :-
-- -- -- -- -- -- -- -- -- -- -- -- --
9
CALIBRATING
9
9
300
-- -- -- -- -- -- -- -- -- -- -- -- --
The display will countdown from 300 to zero in one second steps. If the analyser has
been used recently it may complete automatic calibration in less than 300 seconds
otherwise it will count to zero.
Once the calibration sequence is
complete an audible beep will be heard
and will show the selected fuel on the
following display:-
-- -- -- -- -- -- -- -- -- -- -- -- --
NATURAL GAS
PRESS ENTER KEY
-- -- -- -- -- -- -- -- -- -- -- -- --
Press
This zeros the toxic sensor and sets Oxygen to 20.9%. The next screen is the
MAIN
DISPLAY
of the analyser:-
DATE
. . .
07-08-96
TIME
. . .
12:31:35
INSTABILITY 0
BATTERY %
. .
54
MAIN DISPLAY
NOTE :
It is advisable to repeat Oxygen Calibration every 2 hours for maximum
accuracy.
Use and to change the display.
NETT C
. . . .
0.0
O2 %
. . .
20.9
CO ppm
. . .
0000
EFF (G) %
. . .
0.0
All parameters are detailed in Appendix A - MAIN DISPLAY PARAMETERS.
ENTER