Battery Disposal
Always dispose of depleted batteries using approved disposal methods
that protect the environment
3.
BEFORE USING THE ANALYSER FOR THE FIRST
TIME:
Set the correct time, date etc, after it is switched on and calibrated
– See
USING THE ROTARY DIAL
below. These are stored when the analyser
is switched off.
4. BEFORE USING THE ANALYSER EVERY TIME:
Check the water trap is empty and the particle filter is not dirty:
-
To empty the water trap, unscrew its red stopper and re-screw
once it is empty.
-
To change the filter, remove protective rubber sleeve, pull out the
water trap , remove the water trap’s particle filter from the spigot
and replace. Reconnect the water trap and rubber protective
sleeve.
Remember to always switch on in outside fresh air.
After switch on check that date and time are correct and battery power is
sufficient.
Avoid exposing the instrument to sudden large temperature changes,
and ensure that the unit reads zero for CO and approx. 400 ppm CO
2
in
fresh outside air before use. If necessary re-zero.