TYPE B BOILER OPEN
FLUE:
15 minute test with results
stored every minute
LIMIT = 10ppm
ALARM = 30ppm
TYPE A COOKER:
30 minute test with results
stored every minute
LIMIT = 30ppm
ALARM = 90ppm
TYPE A WATER
HEATER:
5 minute test with results
stored every minute
LIMIT = 10ppm
ALARM = 30ppm
TYPE A SPACE
HEATER:
30 minute test with results
stored every 1 minute
LIMIT = 10ppm
ALARM = 30ppm
Typical ROOM TEST display
ROOM
TEST
1/30
→
Test progress (Test 1 of 30)
INT
40s
→
Test countdown
CO2
1234p
→
Measured CO2
CO
0p
→
Measured CO
Limit
10p
→
CO limit for this test
The user can stop the Room test at any time by pressing PRINT
If not stopped earlier, the Room test will automatically end after the
designated time.
The Room test series is automatically stored in the memory as a log
number.
When completed the log can be printed immediately by pressing
.