ULTRAFLOW® 54 DN15-125
Kamstrup A/S · Technical description · 5512-2464_D1_GB_09.2021
81
8.5.2
Hold function
When the Hold input is activated (high level applied to input), counting stops at the number of pulses counted.
When the Hold signal is removed (low level applied to input), counting restarts.
The counters can also be reset by pressing the right key on the front panel (Reset).
Hold input
Galvanically separated
Input protection
Against reversed polarity
“Open input”
Count (see
Figure 64
8.5.3
Push-button functions
The left push-button shifts between readings/counts of the two flow sensor inputs. In the display, M1 and M2,
respectively indicate the currently displayed flow sensor inputs/counters.
Figure 65. The left push-button
The right push-button resets both counters (M1 and M2).
Figure 66. The right push-button.
8.5.4
Use of Pulse Tester
Pulse Tester can be used as follows:
Standing start/stop of the flow sensor using the integral pulse counters.
Standing start/stop of the flow sensor using the pulse outputs for external test equipment.
With flying start/stop of the flow sensor using the integral counters controlled by external equipment (Sample &
Hold).
With flying start/stop of the flow sensor using the pulse outputs controlled by external equipment (Sample &
Hold).